Q

I am a biologist and have been working with 35S a lot recently. I have been running acrylamide gels containing about 14.8 kBq of radiation per lane (total about 148 kBq per gel). This radiation is not measureable by a Geiger Muller (GM) counter and I have therefore been concerned about the risk of unnoticed contamination and incorporation after accidental contact. I know liquid scintilation is an appropriate monitoring method, but this cannot be performed continuously during an experiment (unlike a GM counter constantly turned on). Are my worries justified, or are these not detectable quantities and potential contaminations (activity way under ALI) "negligible?"

A

Thank you for your question. 148 kBq is a very small activity and, if accidentally ingested, would not cause a harmful effect.

However, 35S is easily detectable with a GM with a "pancake" probe on it. This probe is about 5 percent efficient for the 35S beta so 148 kBq would be roughly 440,000 cpm. Regardless of whether the activity is harmful if ingested, to prevent unnecessary contamination and ingestion, this type of meter should be used for routine surveillance of the work area and of hands.

A liquid scintillation counter is still the gold standard, but you're quite right, it isn't terribly helpful for routine monitoring!

Kelly Classic
Certified Medical Health Physicist

Ask the Experts is posting answers using only SI (the International System of Units) in accordance with international practice. To convert these to traditional units we have prepared a conversion table. You can also view a diagram to help put the radiation information presented in this question and answer in perspective. Explanations of radiation terms can be found here.
Answer posted on 14 September 2009. The information posted on this web page is intended as general reference information only. Specific facts and circumstances may affect the applicability of concepts, materials, and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice. To the best of our knowledge, answers are correct at the time they are posted. Be advised that over time, requirements could change, new data could be made available, and Internet links could change, affecting the correctness of the answers. Answers are the professional opinions of the expert responding to each question; they do not necessarily represent the position of the Health Physics Society.