My test was a typical fluoroscopy diagnostic exam. I was provided the following information about my radiation exposure from the exam:
- The fluoroscopy on time was 2.7 minutes with a dose area product (DAP) of 18.167 Gy-cm2
- Total dose was 142.988 mGy
- The skin surface dose was 0.17 Gy
Are the skin-surface dose and total dose the same thing? Could you please help me to estimate the effective dose based on the DAP?
The skin-surface dose and the machine-reported total dose from the equipment are not necessarily the same. The skin dose can be higher when the distance from the x-ray tube to the skin is less than the distance to the reference point for the system.
With the information provided, I estimate that the effective dose is between 4.5 and 6.8 mSv. Effective dose is a population average—not the dose to an individual patient. The risk from the radiation is either too low to measure or doesn't exist.
Deirdre H. Elder, MS, CHP, CMLSO