Answer to Question #10024 Submitted to "Ask the Experts"
Category: Medical and Dental Equipment and Shielding — Shielding
The following question was answered by an expert in the appropriate field:
I was making maximum entrance skin exposure measurements on a fluoroscopy system. I was using 3.8 cm of aluminum and a sheet of 0.64 cm of lead. My detector was approximately 25 cm (maybe less) from the aluminum/lead. How much would you estimate my exposure rate would be enhanced because of scatter from these phantoms?
These measurements were performed on a Siemens bi-plane vascular lab. This system has a diamentor in the collimator that measures the exposure from the system and feeds this back to the generator to limit the exposure rate. At first this wasn't calibrated but after we accomplished this, Siemens readings didn't agree with mine. Mine were a little higher. I was just wondering if you thought that I would get 25–35% higher readings from scatter?
The geometry you are using will, indeed, result in a significant amount of backscatter. Johns and Cunningham (Table B-1) shows over 30% backscatter (3 mm Al HVL [half-value layer] and 200 cm2 field). You are using a field of over 600 cm2 so the backscatter would be greater than this, i.e., on the order of 43% based on extrapolation of Table B-1 data.
In addition to backscatter from the lead-aluminum block you will also get scatter from the tabletop, i.e., scatter from the lead-aluminum block to the tabletop and then to the detector. This will increase the scatter significantly and probably cancel out any reduction in scatter at the detector due to the 15 to 18 cm distance. I suspect Siemens is doing their measurement with the detector in air, i.e., suspended off the end of the table so any scatter from the table is eliminated.
It is not unusual with today's fluoroscopy systems to get some penetration through the lead in a test like this. The systems usually drive to 120 kVp or higher and maximum mA. The good news is that the grid is removable. I hope the staff members using the equipment realize that removing the grid during a procedure can reduce the patient and staff doses by about 50%. My recommendation would be to start out without the grid and if the images are acceptable then proceed. If the images are of low contrast then insert the grid and continue the procedure. The grid provides minimal benefit in a lot of cases, especially for thinner patients and thinner body parts.
Joel Gray, PhD
Reference
Johns HE, Cunningham JR. The physics of radiology. 4th ed. Springfield: Charles C. Thomas; 1983.